Strain gauges · Measurement

Shunt Calibration Resistor Calculator

Calculate the shunt resistor R_N for a desired strain-gauge calibration signal, exactly and per approximation eq. 8.7, and the strain it simulates.

DMSKeil 2017
01

Inputs

Applies to the balanced bridge with constant excitation; lead resistance between shunt and arm alters the signal slightly at DC and considerably at high carrier frequency.

02

Results

Enter values and run the calculation.

Method

What is calculated?

A high-value resistor across one arm lowers its resistance and unbalances the balanced bridge reproducibly – the whole measuring chain can be checked without mechanical load. Keil's example: 0.5 mV/V on a 120 Ω bridge needs a 60 kΩ shunt per eq. 8.7; exactly 59.94 kΩ.

Equations

R_K = R_N·R/(R_N+R) (Keil Gl. 8.1)

U_M = ½U_B·R/(2R_N+R) (Keil Gl. 8.5)

U_M/U_B ≈ R/(4R_N) ⇒ R_N = U_B·R/(4U_M) (Keil Gl. 8.6, 8.7)

ε_sim = −4·(U_M/U_B)/k

Limits

Assumptions and typical mistake

Applies to the balanced bridge with constant excitation; lead resistance between shunt and arm alters the signal slightly at DC and considerably at high carrier frequency.

Placing the shunt across the whole half bridge instead of one arm, or interpreting the negative sign of the simulated signal as a fault.

Inputs

What you enter – and where the values come from

Arm resistance R (gauge) [Ω]
Resistance of the bridge arm the shunt is placed across – for a self-installed gauge its nominal resistance (120 or 350 Ω), for a finished transducer the resistance of one bridge arm per datasheet. Source: gauge package or transducer datasheet.
Desired calibration signal U_M/U_B [mV/V]
Desired test signal in millivolts of output per volt of excitation (mV/V). A shunt is a known resistor connected temporarily in parallel with one bridge arm to electrically simulate a defined load. Source: your chosen test point near rated output or the shunt-calibration value in the transducer certificate.
Gauge factor k
The gauge factor is the sensitivity of the strain gauge: it states how much the electrical resistance changes when the gauge is stretched (ΔR/R = k·ε). Without it no strain can be computed from an electrical signal. Source: printed on every gauge package or in the manufacturer datasheet, typically 2.0 to 2.1 for constantan foil gauges; valid at room temperature. The same value must be set in the amplifier.
Excitation U_B [V]
The bridge excitation is the DC or AC voltage with which the amplifier supplies the bridge circuit. The signal is proportional to it – which is why signals are usually stated as a ratio in mV/V and the excitation is only needed for the absolute signal voltage. Source: amplifier setting, usually 2.5 V, 5 V or 10 V; the upper limit is set by gauge self-heating (see the excitation calculator).
Context

What this calculator is for

Checking and reproducing the calibration of force transducers and load cells in situ, function checks of remote measuring points, scaling of data-acquisition systems.

What comes next

  1. Enter the arm resistance and the desired calibration signal (e.g. transducer rated output).
  2. Pick the closest E-series resistor to the exact value and back-calculate the actual signal.
  3. Switch the shunt in at the transducer where possible; mind capacitive cable asymmetry at 5 kHz carrier frequency.
Results

How to read the results

  • R_N exact is the target shunt value; choose the nearest standard value and back-calculate the actual signal with it.
  • The simulated strain is the check value for the amplifier reading: if it matches, gauge factor, bridge factor and wiring are consistent.
Result quantities

What each value means

R_N per approximation eq. 8.7 [kΩ]
Approximate value of the test resistor connected in parallel. It is usually sufficient for selecting a component at small test signals up to about 1 mV/V; use the exact value below for accurate calibration.
R_N exact per eq. 8.5 [kΩ]
Shunt resistor per the exact relation; it differs from the approximation by R/2. Use this value for high-accuracy calibration.
Simulated strain (quarter bridge) [µm/m]
Strain the amplifier (in quarter-bridge mode with the entered gauge factor) must display when the shunt is connected. Negative because the parallel resistor mimics compression. If the instrument shows something else, gauge factor, bridge factor or wiring is wrong.
Calibration voltage U_M [mV]
Absolute calibration voltage in millivolts at the entered excitation.
Source

Technical basis

Keil, Dehnungsmessstreifen, 2nd ed. 2017, sec. 8.7.3 shunt calibration, eqs. (8.1)–(8.7), fig. 8.7.

The source supports the equation structure and worked examples; this calculator does not replace calibration of the measuring chain.

FAQ

Frequently asked questions

Does the shunt replace a force calibration?

No, it only reproduces an electrical signal; the signal ↔ quantity relation must be established once with a reference standard.

Why a negative signal?

The parallel resistor reduces the arm resistance like a compression of the gauge.

What accuracy does the shunt need?

At least the target calibration accuracy, typically 0.01 to 0.1 %, with a low temperature coefficient.