- Reading channel X: F_x* [kN]
- The reading of measuring channel X, i.e. the measuring point installed for the force F_x – already converted to force (amplifier scaled to kN). The asterisk means: contaminated by crosstalk, not yet the true force. Source: the reading of the amplifier or data acquisition for this channel under load.
- Reading channel Y: F_y* [kN]
- The reading of measuring channel Y, i.e. the measuring point installed for the force F_y – already converted to force (amplifier scaled to kN). The asterisk means: contaminated by crosstalk, not yet the true force. Source: the reading of the amplifier or data acquisition for this channel under load.
- Reading channel Z: F_z* [kN]
- The reading of measuring channel Z, i.e. the measuring point installed for the force F_z – already converted to force (amplifier scaled to kN). The asterisk means: contaminated by crosstalk, not yet the true force. Source: the reading of the amplifier or data acquisition for this channel under load.
- q_xx (channel X ← F_x)
- The main sensitivity of channel X: reading of the channel divided by the applied force F_x when only F_x acts. If the display was adjusted to the force during calibration this value is 1. Source: calibration with F_x acting alone.
- q_xy (channel X ← F_y)
- The crosstalk coefficient: reading of channel X divided by the applied force F_y when only F_y acts (Keil eqs. 10.79–10.81). A value of 0.1 means: 10 % of the force F_y falsely appears as F_x. Zero for ideal decoupling. Enter as a factor, not in percent. Source: calibration curve with F_y acting alone (Keil Fig. 10.42).
- q_xz (channel X ← F_z)
- The crosstalk coefficient: reading of channel X divided by the applied force F_z when only F_z acts (Keil eqs. 10.79–10.81). A value of 0.1 means: 10 % of the force F_z falsely appears as F_x. Zero for ideal decoupling. Enter as a factor, not in percent. Source: calibration curve with F_z acting alone (Keil Fig. 10.42).
- q_yx (channel Y ← F_x)
- The crosstalk coefficient: reading of channel Y divided by the applied force F_x when only F_x acts (Keil eqs. 10.79–10.81). A value of 0.1 means: 10 % of the force F_x falsely appears as F_y. Zero for ideal decoupling. Enter as a factor, not in percent. Source: calibration curve with F_x acting alone (Keil Fig. 10.42).
- q_yy (channel Y ← F_y)
- The main sensitivity of channel Y: reading of the channel divided by the applied force F_y when only F_y acts. If the display was adjusted to the force during calibration this value is 1. Source: calibration with F_y acting alone.
- q_yz (channel Y ← F_z)
- The crosstalk coefficient: reading of channel Y divided by the applied force F_z when only F_z acts (Keil eqs. 10.79–10.81). A value of 0.1 means: 10 % of the force F_z falsely appears as F_y. Zero for ideal decoupling. Enter as a factor, not in percent. Source: calibration curve with F_z acting alone (Keil Fig. 10.42).
- q_zx (channel Z ← F_x)
- The crosstalk coefficient: reading of channel Z divided by the applied force F_x when only F_x acts (Keil eqs. 10.79–10.81). A value of 0.1 means: 10 % of the force F_x falsely appears as F_z. Zero for ideal decoupling. Enter as a factor, not in percent. Source: calibration curve with F_x acting alone (Keil Fig. 10.42).
- q_zy (channel Z ← F_y)
- The crosstalk coefficient: reading of channel Z divided by the applied force F_y when only F_y acts (Keil eqs. 10.79–10.81). A value of 0.1 means: 10 % of the force F_y falsely appears as F_z. Zero for ideal decoupling. Enter as a factor, not in percent. Source: calibration curve with F_y acting alone (Keil Fig. 10.42).
- q_zz (channel Z ← F_z)
- The main sensitivity of channel Z: reading of the channel divided by the applied force F_z when only F_z acts. If the display was adjusted to the force during calibration this value is 1. Source: calibration with F_z acting alone.